Although most functions of the SUNA V2 can be accomplished though the USB connection, parts of the self-test require that external power be applied. In this case, connect external power and repeat the test. If the problem persists, please contact Customer Support.
SUNA
How does absorption from CDOM affect the SUNA V2 nitrate measurement?
The SUNA V2 determines nitrate concentrations from the shape of the UV absorption curve. The least squares curve fitting algorithm uses calibrated extinction coefficients for nitrate and bromide (strong absorbing species in salt water) to calculate the concentration...
What is “adaptive sampling” and how does it work?
The SUNA V2 contains a 256 channel spectrometer that is programmed to integrate for a specific length of time (usually 300 - 500 ms) while sampling to maximize signal. That is, when the SUNA V2 takes a sample, the spectrograph collects UV light for the length of the...
Should I provide my ISUS or SUNA with a warmup time before recording data?
We recommend updating the ISUS or SUNA's reference spectrum portion of the calibration with the instrument in the same state it will be used in the field. To update the reference spectrum, you will have to run the calibration feature in ISUSCom or UCI using a pure...
What is the common or base environmental sensor suite, and what options are there for other sensors?
The base sensing suite includes a FastCat CTD, a combined fluorescence and turbidity sensor, and a PAR sensor. The current supported list of instruments for Thetis include: CTD: SeaBird FastCat CTD (SBE49) Dissolved oxygen: SeaBird SBE43 or SBE63 Chlorophyll: ...
When running my SUNA V2 on the bench, the data shows that the lamp is turning on and off. Why is this happening?
This is a safety feature of the instrument, whereby the lamp is turned off when its temperature reaches 35°C (95°F) to prevent damage from overheating. The SUNA continues to output dark frames while monitoring the temperature, and will restart the lamp once it has...
I am confused by all these software names. Which software does what?
Sea-Bird’s main software package is called Seasoft©. Seasoft V2 — Seasoft V2 is actually a suite of stand-alone programs. You can install the entire suite or just the desired program(s). Deployment Endurance Calculator — calculates deployment length for moored...
Featured Posts
Ocean Business 2025
We hope to see you at Ocean Business 2025 Overview Ocean Business 2025 is the global meeting place for ocean science and technology professionals. Connect with thousands of the industry’s brightest minds, and share ideas to help define the future of ocean technology....
The Deep SeapHOx™ V2 Moored System
The Deep SeapHOx™ V2: Revolutionizing Moored Ocean Monitoring Welcome to the future of oceanographic research with the Deep SeapHOx™ V2. Designed for long-term deployments in diverse environments, from shallow regions to the deep ocean, this state-of-the-art...
The HydroCAT-EP V2 Multiparameter CTD
The HydroCAT-EP V2 Multiparameter CTD: Precision Insights for Environmental Monitoring Dive into the future of environmental monitoring with Sea-Bird Scientific's cutting-edge HydroCAT-EP V2 (HCEP-V2). This multiparameter CTD offers an impressive dynamic range,...
Science and Technology
Platform