Blog Home » FAQs » Why does the specified range of the ECO BB sensors (expressed in m^-1) not match the maximum output calculated from my calibration sheet (expressed in m^-1 sr^-1)?

Blog Home » FAQs » Why does the specified range of the ECO BB sensors (expressed in m^-1) not match the maximum output calculated from my calibration sheet (expressed in m^-1 sr^-1)?

The back-scattering measurements are a portion of the total beam attenuation coefficient of the water being sampled. The beam attenuation is measured during the back-scattering calibration. Therefore, the range specification on the data sheet for the BB sensor (0 – 3 or 0 – 5 m^-1) refers to the beam attenuation coefficient range of the calibration.

There is a note on the data sheet that further explains the backscattering specification:

*Backscattering specifications are given in beam cp (m^-1) based on the regression of the response of the instrument relative to the beam cp measured at the coincident wavelength using an ac-s spectrophotometer. Scale factors for backscattering incorporate the target weighting function and the solid angle subtended.