Blog Home » FAQs » Can I use a pressure sensor above its rated pressure?

Blog Home » FAQs » Can I use a pressure sensor above its rated pressure?

Digiquartz pressure sensors are used in the SBE 9plus, 53, and 54. The SBE 16plus V2, 16plus-IM V2, 19plus V2, and 26plus can be equipped with either a Druck pressure sensor or a Digiquartz pressure sensor. All other instruments that include pressure use a Druck pressure sensor.

  • The overpressure rating for a Digiquartz (as stated by Paroscientific) is 1.2 * full scale. The sensor will provide data values above 100% of rated full scale; however, Sea-Bird does not calibrate beyond the rated full scale.
  • The overpressure rating for a Druck (as stated by Druck) is 1.5 * full scale. The sensor will provide data values above 100% of rated full scale; however, Sea-Bird does not calibrate beyond the rated full scale.

Note: If you use the instrument above the rated range, you do so at your own risk; the product will not be covered under warranty.

Category: Field Procedures & Deployment