Q&A: Temperature Mismatch
Although they should match perfectly, the two sensors rarely reported the same temperature value throughout their deployment. Furthermore, the 16plusV2 appeared to stop logging, although the batteries still had life upon recovery. Can you identify the problem?
Q&A: Diagnose This Profile
The data shows conductivity from a 19plus deployed in a shallow estuarine river. The red line shows the CTD downcast and the green line shows the upcast. While the upcast data show the expected shape for a CTD profile (low conductivity at the surface, rising conductivity as the CTD descends), the downcast shows an unusual shape in until about 2.6 meters depth, where conductivity appears to suddenly correct itself.
Q&A: Diagnosing Salinity Spikes
Q&A: Diagnosing Salinity Spikes The figure above shows a time series of salinity from a moored CTD. For most of the deployment, salinity showed frequent downward spikes, dropping from an expected range of 18-26 psu to near zero. Can you determine the cause?...
Ocean Sciences 2020 Conference Posters
Missed us at Ocean Sciences 2020? Check out digital copies of our Science Team’s conference posters to learn more about what they are working on: Field Performance of an ISFET Based Profiling pH SensorCharles W. Branham, Vladislav Simontov, Yuichiro Takeshita &...
Q&A: Two Symptoms of the Same Problem
May 2020 Newsletter The data above show two major symptoms of the same problem occurring in different locations on a 19plusV2 profiling CTD. On the left, oxygen appears to drop below zero and is highly erratic during the downcast (light blue). However, once it hits a...
Tech Tip: Troubleshooting a misfiring bottle
Tech Tip: Troubleshooting a misfiring bottle The SBE 32 and SBE 55 water samplers utilize a unique electromagnetic design to reliably capture water. Did you know that the latches are also dishwasher safe? Watch this video to learn how to remove and clean water sampler...
Q&A: Moored Data Excursions
February 2020 Newsletter The data above came from a HydroCAT-EP multiparameter probe moored in an estuary near Savannah, Georgia. The highlighted section shows a sudden change in the temperature and salinity data that lasted for approximately 4 days, and continued...
Q&A: Negative Fluorescence Data
December 2019 Newsletter Fluorescence is the light emitted by a particle after excitation from another light source. Sea-Bird Scientific fluorometers are widely used to measure fluorescence from chlorophyll as a method of measuring primary productivity. The figure...
Q&A: Erratic Profiles
November 2019 Newsletter The data above come from a profiling CTD, deployed in sea conditions that caused approximately 1-meter ship heave. Two things stick out – relatively unchanging salinity data during the downcast, and noisy data during the upcast (shown in light...
Featured Posts
Ocean Optics 2024
We hope to see you at Ocean Optics Conference XXVI 2024 Please join us for our exhibition at the Ocean Optics Conference XXVI in Las Palmas De Gran Canaria, Spain, from October 6-11, 2024. Overview The Ocean Optics Conference attracts a diverse audience of active...
OCEANS 2024
We hope to see you at OCEANS 2024 We are excited to sponsor and exhibit at the upcoming OCEANS Conference & Exposition in Halifax, NS, Canada from September 23 -26. Please come visit our team at booth #213. Overview The OCEANS 2024 Halifax conference is for global...
UG2 Workshop 2024
We hope to see you at UG2 '24 We are excited to sponsor the upcoming 2024 Glider Workshop in Ann Arbor, Michigan, from September 10 - 12, 2024. Overview This workshop will bring together the global underwater glider community to strengthen international collaboration...
Science and Technology
Platform